CR PHANTOM 6×10
Article No.: 11 00218
according to ISO 16371-1, ASTM E2445
Smaller, simpler CR Phantom
for Qualification of CR Systems
Process Control and Long-Term-Stability
Test Phantom for Qualification of Image Plate Scanner Systems (CR)
CR Phantom enables to test all relevant parameters of CR Scanner systems and Imaging Plates (IP) like Basic Spatial Resolution, MTF, Contrast Sensitivity, Blooming, Laser Jitter, Scanner Slippage, Shading, Erasure and more. These tests required and described in detail in standards ASTM E2445 and ISO 16371-1 have to be performed periodically for checking process control and long-term-stability.
The KOWOTEST CR Phantoms exceed these standards by including two Duplex Wire Type IQIs and measuring points for Shading in both axes – portrait and landscape. All required information is mapped on the Imaging Plate with a single X-ray exposure. Phantom must not be rotated to generate information of second axis. This results in higher accuracy and significant time savings.
CR Phantom 6×10 is designed for use of smaller Imaging Plates. Even IPs in weld formats down to 10×24 cm can be used to perform all essential tests.
For Linearity check in landscape direction additional tungsten balls in a graduation of 25 mm are available. The Duplex Wire Type IQIs can be removed to use for other applications.
IQI, Gauge | Test, Check | Deviation | Difference | |
A | T-Target Laser | Jitter, MTF, PMT Non-Linearity | smaller | equal tests feasible |
B | Duplex Wire Type IQI | Basic Spatial Resolution, Unsharpness | optional | D13, D15, D17 to be inserted |
C | BAM Snail | Central Beam Alignment | no | no |
D | Line Pair Gauge | Resolution | missing | test preferably done with Duplex IQI |
E | Shading IQI | Shading | closer | equal tests feasible |
F | Cassette Position Locator | Position of IP | missing | none-essential |
G | Homogeneous Strip Target | Scanner Slippage | smaller | equal tests feasible |
H | Lucite Plate | Carrier Plate | smaller | equal tests feasible |
I | Linearity IQI – cm-/ inch Ruler | Spatial Linearity, Distortion | cm only | equal tests feasible |
I | Tungsten Balls, Ø 1.5 mm | Spatial Linearity, Distortion | additional | for IP width less than 15 cm (6 inch) |
J | Contrast Sensitivity Gauge | Contrast Sensitivity | Cu missing | none-essential, Cu not penetrated at 90 kV |
Dimensions: | 250 x 160 x 16,5 mm | 6 x 10 x 0.65 inch |
Types:
Article No.:
CR Phantom 6×10 – Type I in wooden case
Duplex Wire Type IQIs not included but required (optional)
Test Certificates acc. to ASTM E2445, ISO 16371-1
Declaration of Conformity acc. to EN ISO/ IEC 17050-1/ -2
11 00218
You can find accessories on the leaflet
leaflet German:
leaflet English: CR PHANTOM 6×10